- made from high-quality materials (cylinder: PP / piston: PE-HD, 0.1 ml: LCP)
- PD-Tips II cannot be autoclaved; the adapters of the 25 ml and 50 ml sizes can be autoclaved at 121°C
- in compliance with ISO 8655
- DE-M marked, with batch certificate
BRAND® PD-Tips II can be used also with compatible, third-party dispensing systems. The automatic size detection option is available with:
- BRAND HandyStep® touch
- BRAND HandyStep® electronic
- Gilson® Repetman®
- Rainin AutoRep™ E
In addition, PD-Tips II can be used with:
- BRAND HandyStep® S
- BRAND HandyStep®
- Rainin AutoRep™ S
- Rainin AutoRep™ M
- Eppendorf® Multipette® 4780
- Eppendorf® EDOS® 5221
Non-sterile:
Cat. No. | Capacity | Unit |
---|---|---|
5107706 | 0,50 ml | 100 |
5107708 | 1,25 ml | 100 |
5107712 | 2,50 ml | 100 |
5107713 | 5,00 ml | 100 |
5107714 | 12,5 ml | 100 |
5107715 | 25 ml incl. Adapter | 50 |
5107716 | 50 ml incl. Adapter | 25 |
5107719 | 10 ml | 100 |
5107721 | 1 ml | 100 |
5107723 | 0,10 ml | 100 |
Non-sterile, PD-Tip II Set (20 PD-Tips each in sizes of):
Cat. No. | Capacity | Unit |
---|---|---|
5107730 | 0,5 ml, 1 ml, 1,25 ml, 2,5 ml, 5 ml, 10 ml + 12,5 ml | 1 |
Accuracy table
HandyStep® with PD-Tips II |
Volume range | Nominal volume A* ≤ ± % 10% |
Nominal volume A* ≤ ± % 2% |
Nominal volume CV* ≤ % 10% |
Nominal volume CV* ≤ % 2% |
---|---|---|---|---|---|
0,1 ml | 2-10 μl | 1,6 | 8,0 | 2,0 | 5,0 |
0,5 ml | 10-50 μl | 0,8 | 4,0 | 0,6 | 1,4 |
1 ml | 20-100 μ | 0,8 | 4,0 | 0,4 | 1,0 |
1,25 ml | 25-125 μl | 0,8 | 4,0 | 0,3 | 0,8 |
10 ml | 200-1000 μl | 0,4 | 1,5 | 0,2 | 0,5 |
12,5 ml | 250-1250 μl | 0,3 | 1,5 | 0,2 | 0,4 |
2,5 ml | 50-250 μl | 0,7 | 3,5 | 0,2 | 0,8 |
25 ml | 500-2500 μl | 0,3 | 1,5 | 0,2 | 0,4 |
5,0 ml | 100-500 μl | 0,5 | 2,5 | 0,2 | 0,6 |
50 ml | 1000-5000 μl | 0,3 | 1,5 | 0,15 | 0,4 |
* Calibrated to deliver (TD, Ex). Error limits according to the nominal capacity (= max. volume) indicated on the instrument, obtained with instrument and distilled water at equilibrium with ambient temperature at 20 °C, and with smooth operation